Vascular assessment using PeriFlux 5000 - toe/ankle pressure, ABI/TBI, PVR, tcpo2, SPP - all in one | Perimed
Home Clinical Vascular Assessments

Vascular assessments PeriFlux System 5000

Print

Take your diagnosis one step further!

As a general recommendation all stages of Peripheral Arterial Disease (PAD) should be verified and confirmed using objective tests such as ankle/brachial index (ABI), toe pressures, tcpO2 and microcirculatory evaluations.(TASC II)

Combining different technologies in the same modular solution and software, the Vascular Lab - PeriFlux System 5000 gives you the tools to perform most of these objective tests using the same equipment. This will streamline your clinical workflow and radically simplify learning.
 

All diabetic patients with an ulceration should be evaluated for PAD using objective tests.
Recommendation 18. (TASC II)

Critical Limb Ischemia is a clinical diagnosis but should be supported by objective tests.
Recommendation 19. (TASC II)

 

One instrument, one software -
several objective tests for precise and reliable diagnosis of PAD

 

Test 

 Function Unit

PSW ExM

 Guidelines

laser Doppler
(PF5010)

Heat
(PF5020) 

tcpO2
(PF5040)

Pressure
(PF5050)

Toe/ankle pressure, TBI/ABI
» Read more

X     X X

ABI values below 0.9 are indicative of PAD. Values above 1.40 should always be confirmed with toe pressure or other techniques. Diabetics and other patients with suspected calcified vessels should always be tested using toe pressures. Toe pressures below 0.7 are abnormal. (TASC II)

PVR
» Read more

 

 

 

X X

Qualitative assessment relying on the interpretation of the graphical output and the location of the PVR measurement. (TASC II) 

tcpO2
» Read more

 

 

X

 

X

Values below 30 mmHg are considered critical. (TASC II)

Heat controlled laser Doppler
» Read more

X X

 

 

X

Values above 20 perfusion units during heating and an increase greater than 150% compared to the baseline value predict wound healing.
(Wound care practice, edited by Paul Sheffield et al. Best Publishing Company 2004, p 117-156)

SPP
» Read more

X

 

 

X X  

ABI= ankle-brachial index, TBI=toe-brachial index, PVR = pulse volume recording, tcpO2=transcutaneous oxygen, SPP=skin perfusion pressure 

Learn more about the Vascular Lab - PeriFlux System 5000 and vascular assessments
Clinical study - Use of Transcutaneous Oximetry and Laser Doppler with Local Heat Provocation to Assess Patients with Problem Wounds