PeriCam PSI System Technical Specifications
|
|
|
| Measurement Principle |
LASCA (LAser Speckle Contrast Analysis) |
| Image Size: |
Normal Resolution model: ~5.9 x 5.9 cm – 15 x 15 cm
High Resolution model: ~25 x 34 mm |
| Image Acquisition Rate: |
50 Hz: 94, 47, 23, 18, 8, 3, 2, 1, 0.5, 0.2 images per second
60 Hz: 112, 55, 28, 22 10, 3.5, 2, 1, 0.5, 0.2 images per second |
| Image Resolution: |
Maximum 1386 x 1034 measurement points |
| Camera Resolution: |
Measurement Camera: 1388 x 1038 pixels
Documentation Camera: Color, 752 x 580 pixels |
| Pixel/Measurement Point Resolution: |
Normal Resolution model: ~ 10 000 pixels per cm2;
100 μm/pixel (at closest working distance)
High Resolution model: ~250 000 pixels per cm2;
20 μm/pixel |
| Laser Specifications: |
Measurement laser: 785 nm, 70 mW,
Class 1 per IEC 60825-1:2001 -
Safe to use without eye protection
Area indicator laser: 660 nm, 5mW,
Class 1 per IEC 60825-1:2001 -
Safe to use without eye protection |
| Dimensions and Weight: |
Scanner head: 22 x 15 x 20 cm, 2.0 kg |
Due to Perimed’s commitment to continuous improvement of our products, all specifications are subject to change without notice.
|